FastQC reports in the overrepresented sequences module a warning for all R2 FASTQ's corresponding to a stretch of >50 G's, amounting to <0.5% and <0.3% of the whole library before and after removing of adapters (TrueSeq Nano), respectively.
Our partner that performed the labwork advised not to concern about this, idicating that it is a known artifact of NovaSeq runs and that their quality checks did not pick up on this.
I would like to get some advice on whether or not I should deal with this warning (i.e. add this GGG...GGG to the adapter file passed to trimmomatic), or should I just leave it as it is and proceed with the alignments.
Hi serpalma.v ,
Novaseq has the same technique as the NextSeq. The 'G' is identified by detecting no signal (here). Thus, if you run out of your template, the machine reports a G-stretch. This stretch will have a very low sequence quality and trimmed using low-quality tail trimming.
Thus, I'd get rid of the poly-G.
Cheers,
Michael
Thanks Michael,
is there a tool that you would recomend to do the job? I am not sure if trimmomatic is the right tool for this.
Thanks
I've worked with bbduk.sh from BBMAP and with cutadept. The first is quite fast and flexible the later is better suited for PE-reads (IMHO).