Hello,
I have two samples that were sequenced using pacbio technology. These samples are different in terms of their shearing methods. I also have SMRTLink installed through command line interface.
I want to run some primary analysis to compare these samples, for example coverage bias, alignment summary metrics, and any thing else that can help with understanding differences between these samples.
I have unaligned bam file, pbi and xml files for each sample. I looked into pacbio guid but I couldn't find anything that helps me with these metrics.
Is there any tool I can use within SMRTLink that can do these analysis?
Thanks
1 answer
You're looking for the Mapping Analysis application - cromwell.workflows.pb_align_ccs on page 103 of the SMRTLink 10.1 docs you linked.
Running pbcromwell run pb_align_ccs ... will allow you to run the GUI mapping analysis on the command line.
or pbcromwell run pb_align_subreads... if you have subread data.
You can also just use pbmm2 that's included w/ the SMRTlink install to do the mapping manually if you are so inclined.
https://github.com/PacificBiosciences/pbmm2